Noise30 db
-
Weight1.5 kg
-
Warranty2 kg
Point 2 AutoCPAP
Within a certain pressure range the device automatically controls the therapy pressure depending on the occurrence of respiratory events. In sleep phases with no events the therapy pressure is lowered. Starting obstructive events are treated effectively with a slight pressure increase. For a high therapy comfort the therapy pressure should always be only as high as necessary and as low as possible.
The device can recognize the following respiratory events:
- Obstructive apnea
- Hypopnea
- Central apnea
- Snoring
- Normal respiration
- Increased airway resistance
The identification of the different respiratory events is achieved via 3 channels:
- Measurement of any constriction of the respiratory tract
- Detection of flow limitations by measuring the flow
- Detection of snoring with the help of sound detection
The combined analysis of all results facilitates a very precise differentiation between obstructive and central events. The device adjusts the therapy pressure according to the detected events as described below:
- Increase of pressure for obstructive events
- Decrease of pressure for central events and normal respiration
In addition to normal APAP and CPAP mode the device offers the FLEXLINE mode, a pressure increase which is proportionate to the individual respiration flow during inspiration and a pressure decrease during expiration
- 3 FLEX-Levels separately for in- and expiration
- Data management on SD-Card
- Remarkable low noise level
- Battery operation possible
- Tidal volume display
Specifications:
Range | Steps | Point2 Auto CPAP |
|
Pressure | 4 hPa – 20 hPa | 0.5 hPa | |
P-Ramp | 4 hPa – Pressure - IPAP | 0.5 hPa | |
P-Min | 4 hPa – P–Start | 0.5 hPa | |
P-Max | P–Start – 20 hPa | 0.5 hPa | |
P-Start | P-Min - P-Max | 0.5 hPa | |
I-FLEX | 0 - 3 | 1 step | |
E-FLEX | 0 - 3 | 1 step | |
IPAP | EPAP – 20 hPa | 0.5 hPa | |
EPAP | 4 hPa – IPAP | 0.5 hPa | |
I-Slope | 1- 3 | 1 step | |
E-Slope | 1- 3 | 1 step | |
Frequency | 5 – 30 bpm | 1 bpm | |
I:E | 20 - 80 % | 1 % | |
Delay time | 0 - 20 s. | 1 s. | |
S- Frequence |
On (10 bpm)
Off
|
||
Mask | Nose/Nose- Mouth/Pillow | ||
Start-Stop Automatic Calibration | Start via ENTER | ||
Automatic | Start-Stop/Start/Off | 10 min. | |
Mask Test | 0 - 90 s. | 5 s. | |
Ramp | 0 – 45 min. | 5 min. | |
Display Vt | On / Off | ||
Bacterial Filter | Yes / No | ||
Brightness | 0 - 100 % | 10 % | |
Language | deu, eng, ell, spa, fra, ita, nld, tur, plk, por, cze | ||
P- Unit | Hpa/mbar/sm H2O | ||
Power | Counter | ||
Blower | Counter | ||
Filter | Counter | ||
Therapy | Counter | ||
Date | Day-Month-Year | ||
Time | Hour-Minute | ||
Waketime | Hour-Minute |
Technical Data:
- Dimensions:- 170 mm x 220 mm x 95 mm (W x L x H)
- Mass:- 1.51 kg
- Power supply:- 100 V - 240 V AC, 50 - 60 Hz
- Power consumption- < 17 W
- DC operation - 24 V DC / 2.1 A (max.)
- Battery operation - 2 x 11.1 V / 2200 mAh
- Pressure range - 4 - 20 hPa (mbar)
- Pressure variation- < 0.5 hPa without humidifier; < 0,9 hPa with humidifier
- Noise level - < 30 dB(A) at 10 hPa
Functions:
- Counters- stand-by time, turbine running time, therapy time, filter age
- Permanent display- pressure, bargraph, Date, time, softstart ramp, alarm clock tidal volume (Vt), battery charging symbol, pressure unit (hPa, mbar, cmH 2O)
- Comfort functions- start-stopp-automatik, softstart ramp, leakage test, mask calibration, flex pressure adjustment I-Flex / E-Flex, automatic start after power failure, date-, time and wake-up function, mask alarm.
SD-Card
Therapy data | Internal | SD card |
Central apnea, obstructive apnea, mixed apnea, hypopnea | saved | saved |
Airway constriction, snoring, hyperventilation | saved | saved |
Leakage | saved | saved |
Base Pressure | Only saved, when I- Flex or E- Flex are selected, but not active | Only saved, when I- Flex or E- Flex are selected, but not active |
Therapy Pressure (low resolution). No FLEX pressure changes visible | saved | not saved |
Therapy Pressure (high resolution), respiration flow, relative respiratory volume. FLEX pressure changes visible | not saved | saved |
Storage Time | </= 30 days | </= 2 years |
Aquapoint 2 (Not Included)
- Dimensions-(w x l x h) 100 mm x 145 mm x 109 mm
- Mass approx.- 280 g (without water)
- Power supply- 24 V DC
- Power consumption- 0 - 22 W
- Heating levels- 5 stages + off
- Max. filling capacity- 220 ml
Powerpackpoint 2 (Not included)
- Dimensions- (w x l x h) 160 mm x 45 mm x 70 mm per battery
- Mass approx.- 300 g per battery
- Power supply- 11.1 V DC per battery
- Capacity- 2200 mAh per battery
Filtersystempoint 2
- Dimensions- (w x l x h) 153 mm x 45 mm x 31 mm
- Mass approx.- 50 g
- Coarse filter- Polyurethan on polyester based 80 ppi (pores per inch), 30 kg/m³ (material density)
- Fine filter- Polypropylen Precipitation of particles ≥ 1 µm: ≥ 89% at ω ≤ 1 m/s (particle velocity) ≥ 99% at ω ≤ 0,2 m/s (particle velocity)